Getting Started with the CoaguChek XS System
Return to CoaguCheck XS Main Page​


​                                                                  




















Cleaning/Disinfecting the Exterior

Do not use sprays of any sort.
Ensure that the swab or cloth is only damp, not wet.
Use only the following items for cleaning/disinfecting the CoaguCheck XS meter 
    housing for a contact time of >1 minute:
        70% isopropyl alcohol
        10% Sodium hypochlorite solution (1 part bleach to 9 parts deionized water, 
             made fresh every 24 hours)
        Note: Do not use any other disinfectants/cleaning solutions on the meter 
             housing.

Clean/Disinfect the exterior: With the meter powered off, wipe the meter's exterior 
    clean. Do not let liquid accumulate near any opening. Make sure that no liquid 
    enters the meter.

Dry the exterior: With a lint-free tissue, dry the meter. Wipe away residual moisture
    and fluids after cleaning the housing. Allow wiped areas to dry for at least 10
    minutes before performing a test.

Cleaning/Disinfecting the Test Strip Guide: Use only 70% isopropyl alcohol or 
    10% bleach solution to clean the CoaguChek test strip guide. Do not use any 
    other cleaning/disinfecting solutions on the test strip guide. Use of other 
    cleaning/disinfecting solutions could result in damage to the meter.
        1. Open the cover. With the meter powered off, use your thumbnail to open the
             cover of the test strip guide by pressing its front edge upward. Move the
             safely away from the meter. Then rinse the cover with water or wipe it clean.
        2.  Clean/Disinfect the Test Strip Guide. Hold the meter upright with the test
             strip guide facing down. Clean the easily accessible areas with a cotton
             swab. Ensure the swab is only damp, not wet. Apply cleaning agent for a
             contact time of less >1 minute. Wipe away residual moisture and fluids. Do 
             not insert any objects into the test strip guide. Doing so could damage the 
             electrical contacts behind the test strip guide.
        3.  Allow to dry. With the cover off, allow the test strip guide to dry for at least
             10 minutes before re-attaching the test strip guide cover and testing again.
        4. Close the cover. Close the cover, and make sure it snaps into place.